command: name: "load" summary: "Configuration of the current load setting" query: syntax: "load?" replies: ok: "load:," configuration: syntax: "load=," replies: ok: "ok" error: "error:" parameters:

  • name: "CtrlId" meaning: "Id of load value" type: "unsigned short int" values:
    • code: 4 meaning: "Pedal Force in Newtons"
    • code: 5 meaning: "Power in Watts"
    • code: 6 meaning: "Inclination in %"
    • code: 255 meaning: "will be sent on request"
  • name: "Val" type: "float"

load

Configuration of load value (the same like the manual control mode)

Query command and replies

load? 🡺 load:<CtrlId>,<Val>

If no manually controlled ergometry has been set (e.g., during a stage-controlled ergometry), the value 255 is rendered as <CtrlId>. In this case, no further value will be sent for <Val>.

Configuration command and replies

load=<CtrlId>,<Val> 🡺 ok or error:<message>

(in slave mode only)

Parameters

  • <CtrlId> — Identifier of the controlled load value.
    • 4 — Pedal Force in Newtons
    • 5 — Power in Watts
    • 6 — Inclination in %
    • 255 — will be sent on request (see above)
  • <Val> — Force value in dependency of CtrlId in N, W or %
    • Co-domains:
      • Pedal Force 0..1500 Newtons
      • Power 0..3000 Watts
      • Inclination –25..25 %

Call the load command before starting of ergometry because of initializing of the load type and load value! So you remove also monitoring settings, start conditions and cancel contitions. During the ergometry you must not change load type!

An example in chapter 3.4 demonstrates the application of this command.

Notes

  • Version 4
  • Supported in Release 5.0