Does installing OOS void the warranty?

Installing with a second card means opening the scope, and that voids the warranty. The USB install does not open the scope, but it changes the card, and nothing here promises that your warranty survives that. See the Disclaimer.

Which models does OOS run on?

A DHO924S (the developer's unit) and, by a user report, a DHO804. Models of the same hardware variant differ only in their licensed bandwidth and count as verified. The other DHO800/DHO900 models are prepared but have never run OOS. See Supported models and features.

Can I go back to the stock system?

Yes. OOS runs next to the stock system on the same card. In one-shot mode, the next power-on starts stock. See Going back to stock.

Does OOS change my calibration?

OOS never writes the FPGA flash, /rigol/data, or the uboot, trust and parameter partitions. Self-calibration writes its results to /oos/cal; the factory files are never changed. A mistake while writing the card can still destroy the calibration stored on it, so back up the whole card first (backup).

Will my option licences keep working?

Option licences of the stock system have been seen to stop working after running OOS, and had to be entered again. Keep your licence keys at hand.

Are measurements with OOS accurate?

OOS is not a calibrated or certified measuring instrument. Readings can be wrong without any visible sign. Do not use it for safety-relevant, medical, legal, commercial or compliance measurements.

Is it safe to put the scope on my network?

Only on a trusted bench network, never directly on the internet. SSH is the only remote shell. Apart from SSH, OOS has no authentication: SCPI, VXI-11, the web remote and MCP are open to everyone who can reach the scope. Updates require a signature from this project's release key.

Why did the stock system start instead of OOS?

By default OOS is one-shot: every start of OOS restores the stock boot. Set System > Boot mode to Persist to keep OOS. See First start and choosing how OOS starts and Troubleshooting.

How do I update OOS?

Open http://<scope>:8080/update and choose the .oosupd file of a release. See Updates.

Can I try OOS without a scope?

Yes, with simulated signals. See Try it without a scope.